LaMotte LTC3000wi Turbidity & Chlorine Lab Meter User Manual

Page 66

Advertising
background image

3. Press

ENTER

to select

Free

Chlorine.

Free Chlorine (L)

Scan Blank

Scan Free Chlorine

12:00:00 001/500

4. Rinse a clean tube (0290)

three times with the sample.

Fill the tube to the 10 mL line

with the sample. Cap the

tube. Dry the tube with a lint-

free cloth.

5. Open the meter lid. Insert the

tube into the chamber. Align

the index line on the tube with

the index arrow on the meter.

Close the lid. Press

ENTER

to

scan the blank. The screen

will display

Blank Done for

about 1 second and then

return to the

Free Chlorine

menu.

Free Chlorine (L)

Scan Blank

Scan Free Chlorine

12:00:00 001/500

6. Remove the tube from the

meter. Add 5 drops of DPD

1A Free Chlorine reagent

(P-6740) and add 5 drops of

DPD 1B Free Chlorine reagent

(P-6741). Cap and invert to

mix. The solution will be pink

if chlorine is present. Read

within 30 seconds.

DPD

1B

DPD

1A

66

Chlorine Analysis

Advertising