LaMotte LTC3000wi Turbidity & Chlorine Lab Meter User Manual

Page 42

Advertising
background image

5. Rinse a clean tube (0290)

three times with the blank.

If samples are expected to

read below 1 NTU the meter

should be blanked with a 0

NTU Primary Standard or

prepared turbidity-free (<0.1

NTU) water. For the most

accurate results, use the

same tube for the blank and

the sample.

l

6. Fill the tube to the fi ll line with

the blank. Pour the blank

down the inside of the tube to

avoid creating bubbles. Cap

the tube.

7. Wipe the tube thoroughly with

a lint-free cloth.

8. Open the meter lid. Insert the

tube into the chamber. Align

the index line on the tube with

the index arrow on the meter.

Close the lid.

42

Turbidity Calibration

Advertising
This manual is related to the following products: