LaMotte LTC3000wi Turbidity & Chlorine Lab Meter User Manual

Page 44

Advertising
background image

14. Press

ENTER

to scan the

standard. The screen will

display

Reading for about 1

second. The result will appear

on the screen.

Turbidity WB (F)

0.99

NTU

Scan Blank

Scan Sample

12:00:00 001/500

15. Press

to scroll to

Calibrate.

Turbidity WB (F)

0.99

NTU

Scan Sample

Calibrate

12:00:00 001/500

16. Press

ENTER

to select

Calibrate. A reverse font

(dark background with light

characters) will appear to

indicate that the reading can

be adjusted.

Turbidity WB (F)

0.99

NTU

Scan Sample

Calibrate

12:00:00 001/500

17. Press

or

to

scroll to the concentration

of the standard, 1.00 in the

example. Note: The allowable

adjustment is ±20%.

Turbidity WB (F)

1.00

NTU

Scan Sample

Calibrate

12:00:00 001/500

18. Press

ENTER

to select

Calibrate. Two menu

choices will be offered,

Set

Calibration and Factory

Setting.

Calibrate Menu

1.00

NTU

Set Calibration

Factory Setting

12:00:00 001/500

44

Turbidity Calibration

Advertising
This manual is related to the following products: