9. Press

to scroll to

Calibrate.

Color

110

cu

Scan Sample

Calibrate

12:00:00 001/500

10. Press

ENTER

to select to

Calibrate. A reverse font

(dark background with light

characters) will appear to

indicate that the reading can

be adjusted.

Color

110

cu

Scan Blank

Calibrate

12:00:00 001/500

11. Press

or

to

scroll to the concentration

of the standard, 100 in this

example. Note: The allowable

adjustment is ±25%.

Color

110

cu

Scan Blank

Calibrate

12:00:00 001/500

12. Press

ENTER

to select to

Calibrate. Two menu choices

will be offered, Set Calibration

and Factory Settings.

Calibrate Menu

110

cu

Set Calibration

Factory Setting

12:00:00 001/500

13. Press

ENTER

to select

Set

Calibration. and save the

calibration. Press

or

to scroll to and select

Factory Setting to revert to

the factory calibration. The

meter will momentarily display

Storing... and return to the

Color menu. The calibration

has now been saved and the

meter can be used for testing.

Color

Scan Blank

Scan Sample

12:00:00 001/500

74

Color Calibration