Calibration & analysis – LaMotte LTC3000wi Turbidity & Chlorine Lab Meter User Manual

Page 40

Advertising
background image

CALIBRATION & ANALYSIS

TURBIDITY

CALIBRATION

Turbidity Standards

Only use AMCO or formazin standards with the LTC3000we/wi.

StablCal® standards below 50 NTU should not be used to calibrate

the LTC3000we/wi. The diluent used in the StablCal® standards has

a different refractive index than traditional formazin standards and will

affect the results. The concentration of the calibration standard should

be similar to the expected concentration of sample that will be tested.

The following standards are available from LaMotte Company:
1480 0 NTU/FNU Standard, 60 mL (EPA or ISO)
1450 1 NTU Standard, 60 mL (EPA)
1453 1 FNU Standard, 60 mL (ISO)
1451 10 NTU Standard, 60 mL (EPA)
1454 10 FNU Standard, 60 mL (ISO)
1452 100 NTU Standard, 60 mL (EPA)
1455 100 FNU Standard, 60 mL (ISO)
Turbidity Calibration Procedure

The default units are NTU or FNU and the default calibration curve is

formazin as indicated by (F) in the Menu bar. A LTC3000we, which uses

NTU will be used in the following examples. For the most accurate

results, a user calibration should be performed. The Japan Standard

calibration mode, as indicated by (J) in the Menu bar, should be used

only with Japanese Polystyrene Standards (0-100 NTU). To change the

settings see the Set Up Instructions on page 21.
For the most accurate results, perform a calibration over the smallest

range possible.

Use a calibration standard that, along with the blank,

brackets the range of the samples that will be tested. For example, if

the samples that are to be tested are expected to be below 1 NTU, more

accurate results will be obtained by calibration with a blank and a 1 NTU

standard as opposed to a blank and a 100 NTU standard.
The meter has fi ve measuring ranges:
0 – 11 NTU/FTU

11 – 110 NTU/FTU

110-300 NTU/FTU

300-600 NTU/FTU

600-4000 NTU/FTU

Each range can be calibrated with one point per range. (Six points

total - a blank plus one point in each of the fi ve ranges - if each range

is calibrated.) New calibration points will replace old calibration points

independently for each range. If one range is recalibrated, the meter will

retain the old calibration data for the other ranges. It is recommended

that the meter be calibrated for each range that will be used. The value

40

Turbidity Calibration

Advertising
This manual is related to the following products: