LaMotte LTC3000wi Turbidity & Chlorine Lab Meter User Manual

Page 52

Advertising
background image

3. Remove

the

fi lter from the

syringe. Remove the plunger

from the barrel. (This step is

required to prevent rupturing

the fi lter by the vacuum that

would be created when the

plunger is removed.)

4. Replace

the

fi lter and repeat

step 2 for a second rinse

of the syringe and storage

container.

5. Remove

the

fi lter from the

syringe. Remove the plunger

from the barrel. Replace

the fi lter and fi ll the syringe

with approximately 50 mL

of deionized water. Filter

the water into the storage

container and save this

turbidity-free water.

6. Repeat Step 5 until the

desired amount of turbidity-

free water has been collected.

TESTING TIPS

1. Samples should be collected in a clean glass or polyethylene

container.

2. Samples should be analyzed as soon as possible after collection.
3. Gently mix sample by inverting before taking a reading but avoid

introducing air bubbles.

4. For the most precise results, follow the recommended procedure

52

Turbidity Analysis