LaMotte LTC3000wi Turbidity & Chlorine Lab Meter User Manual

Page 65

Advertising
background image

9. Press

ENTER

to select

Scan

Total Chlorine. The screen

will display

Reading... for

about 1 second. The result

will appear on the screen.

Record the result as Total

Chlorine.

Total Chlorine (T)

1.25

ppm

Scan Free Chlorine

Scan Total Chlorine

12:00:00 001/500

*Warning: Reagents marked with an * are considered to be potential

health hazards. To view or print a Material Safety Data Sheet (MSDS)

for these reagents go to www.lamotte.com. To obtain a printed copy,

contact LaMotte by e-mail, phone or fax.
NOTE: For the most accurate results, samples over 6 ppm chlorine

should be diluted with chlorine demand free water and re-tested.
NOTE: The meter will remember the last scanned blank reading. It is not

necessary to scan a blank each time the test is performed. To use the

previous blank reading, instead of scanning a new one, scroll to Scan

Sample and proceed. For the most accurate results, the meter should

be blanked before each test and the same tube should be used for the

blank and the reacted sample.

CHLORINE: ANALYSIS - DPD LIQUID REAGENTS

Free Chlorine, Combine Chlorine and Total Chlorine

The default units are ppm and the default calibration curve is for DPD

Tablet reagents. For the most accurate results, a user calibration should

be performed. The letter

(L) in the menu bar indicates that the meter is

in the liquid DPD reagent mode. To use tablet DPD reagents, see the Set

Up instructions.

1. Press and briefl y hold

to turn the meter on. The

LaMotte logo screen will

appear for about 3 seconds

and the

Main Menu will

appear.

Main Menu

Measure

Data Logging
Options
Run PC Link

12:00:00 001/500

2. Press

ENTER

to select

Measure. Press

to

scroll to

Free Chlorine.

Measure Menu

Turbidity - No Blank
Turbidity - With Blank

Free Chlorine

Total Chlorine

12:00:00 001/500

65

Chlorine Analysis

Advertising
This manual is related to the following products: