LaMotte LTC3000wi Turbidity & Chlorine Lab Meter User Manual

Page 14

Advertising
background image

Japanese Water Works requirements. The Japanese polystyrene

standards can only be purchased in Japan. Formazin, AMCO and

Stablcal

®

standards cannot be used with this calibration.

TURBIDITY UNITS

Traditionally, turbidimeters designed for use in the United State were

made to the specifi cations of EPA Method 180.1. This method defi ned

the NTU, nephelometric turbidity unit, as a unit to measure turbidity

in the range of 0 – 40 NTU using a nephelometer. According to the

EPA a nephelometer was a turbidimeter that measured turbidity with a

90° detector. Also, if the turbidity was greater than 40 NTU, a dilution

was necessary to bring the sample into the 0 – 40 NTU range. Today,

many turbidimeters have additional detectors which increase the range

of the turbidity measurement, eliminate interferences and generally

improve the performance. Currently, many turbidimeters are capable of

measuring above 40 NTU by using detectors other than a 90° detector.

Even though they may use a 180° detector to measure the light that is

attenuated by high turbidity samples they may continue to report the

results as NTU.
Recently there has been an effort to use the units of turbidity

measurements to indicate which type of detector and light source was

used. For EPA compliant meters, measurements made with a 90° degree

detector and an incandescent white light source are reported as NTU.

When an attenuation measurement is made with a 180°detector, using

the same meter, the results are reported as AU, attenuation units. ISO

Method 7027, which specifi es a 860 nm light source, also uses two

turbidity units. When the 90° degree detector is used, the results are

reported as FNU, formazin nephelometric units. When an attenuation

measurement made with a 180°detector, the results are reported as

FAU, formazin attenuation units. It should be noted that all units are

numerically equivalent if the meters are calibrated to formazin and

that the units only designate which detector was used to make the

measurement. For example, 1 NTU = 1 AU = 1FNU = 1FAU.

14

Advertising
This manual is related to the following products: