LaMotte LTC3000wi Turbidity & Chlorine Lab Meter User Manual

Page 67

Advertising
background image

7. Immediately insert the tube

into the meter. Align the index

line on the tube with the index

arrow on the meter. Close the

lid.

8. Press

ENTER

to scan the

sample. The screen will

display

Reading... for about 1

second. The result will appear

on the screen. Record the

result as Free Chlorine.

Free Chlorine (L)

0.99

ppm

Scan Free Chlorine

Scan Total Chlorine

12:00:00 001/500

9. Remove the tube from the

meter. Add 5 drops of *DPD

3 Total Chlorine Reagent

(P-6741). Cap and invert to

mix. An increase in color

represents Combined

Chlorine.

DPD

3

NOTE: For wastewater samples, Standard Methods for the Examination

of Water and Wastewater recommends waiting 2 minutes for full color

development when testing total chlorine.

10. Open the meter lid. Insert the

tube into the chamber. Align

the index line on the tube with

the index arrow on the meter.

Close the lid.

67

Chlorine Analysis

Advertising
This manual is related to the following products: