8. Press

ENTER

to select

Scan

Sample. The screen will

display

Reading for about 1

second. The result will appear

on the screen.

Turbidity NB (F)

10.22

NTU

Scan Blank

Scan Sample

12:00:00 001/500

ANALYSIS WITH BLANKING PROCEDURE

To obtain the most accurate results the meter should be blanked before

measuring a sample. The blanking step is not as critical for samples

above 10 NTU. The meter should always be blanked before reading

samples below 10 NTU.

1. Press and briefl y hold

to turn the meter on. The

LaMotte logo screen will

appear for about 3 seconds

and the

Main Menu will

appear.

Main Menu

Measure

Data Logging
Options
Run PC Link

12:00:00 001/500

2. Press

ENTER

to select

Measure.

Measure Menu

Turbidity - No Blank

Turbidity - With Blank
Free Chlorine
Total Chlorine

12:00:00 001/500

3. Press

to scroll to

Turbidity - With Blank.

Measure Menu

Turbidity - No Blank

Turbidity - With Blank

Free Chlorine
Total Chlorine

12:00:00 001/500

47

Turbidity Analysis