PROCEDURE

Use universal sample holder.

1.

Press and hold

ON button until spectrophotometer turns on.

2.

Scroll to and select PROGRAMMED TESTS.

3.

Scroll to and select ALL TESTS (or another sequence containing 31
Cu-Cuprizone

) from TESTING MENU.

4.

Scroll to and select 31 Cu-Cuprizone from menu.

5.

Rinse a tube (0290) with sample water. Fill to the 10 mL line with sample.

6.

Insert the tube into chamber, close lid and select SCAN BLANK.

7.

Remove tube from Spectro and add 5 drops of Copper A (P-6367). Cap
and mix.

8.

Add 5 drops of *Copper B (P-6368). Cap and mix.

9.

Wait 5 minutes. Mix.

10.

Insert tube into chamber, close lid and select SCAN SAMPLE. Record
result.

11.

Press

OFF button to turn the spectrophotometer off or press EXIT button

to exit to a previous menu or make another menu selection.

þNOTES: For best possible results, a reagent blank should be determined to

account for any contribution to the test result by the reagent system. To
determine the reagent blank, follow the above test procedure to scan a
distilled or deionized water blank. Then follow the above procedure to
perform the test on a distilled or deionized water sample. This test result is
the reagent blank. Subtract the reagent blank from all subsequent test results
of unknown samples. It is necessary to determine the reagent blank only
when a new lot number of reagents are obtained.

The reaction may stain the tubes. Scrub tubes thoroughly after each use.