Chlorine, free – udv – LaMotte SMART Spectro Spectrophotometer User Manual

Page 88

Advertising
background image

CHLORINE, FREE – UDV

DPD METHOD–UNIT DOSE VIALS • CODE 4311-H

QUANTITY

CONTENTS

CODE

1

*Free Chlorine Unit Dose Vials, 10 pouches

4311-H

Equipment needed but not supplied:

STANDARD ACCESSORY PACKAGE • CODE 1961

1

Package of 3 Vials (empty)

0156

1

Syringe, 3 mL, plastic

1184

1-

Or:

ADVANCED ACCESSORY PACKAGE • CODE 1962

1

Pipettor

30528

1

Pipet Tips (0-5 mL)

30695

1

Cuvette Rack

31695

1

Package of 3 Vials (empty)

0156

1

Foil Storage Bag

9467

*WARNING: Reagents marked with a * are considered to be potential health hazards. To view
or print a Material Safety Data Sheet (MSDS) for these reagents see MSDS CD or
www.lamotte.com. To obtain a printed copy, contact LaMotte by e-mail, phone or fax.

All water for cities and communities must be sanitized; even waters that come
from clean sources, protected watersheds, reservoirs, and deep wells, are
commonly sanitized to assure safety. Chlorine is the most commonly used
sanitizer for several reasons: it is effective against a wide range of
microorganisms, the cost is low, and the methods of applying it have been well
developed. If an adequate concentration of chlorine is present in the water for
a few minutes, disease producing bacteria will be destroyed. A number of
conditions affect the sanitizing action of chlorine. In municipal systems these
can be controlled so that if chlorine is detectable, it can be assumed that
bacteria have been killed. The factors that influence the rate of kill are
temperature, pH, presence of other materials that react with chlorine, time,
and the concentrations of the various chlorine combinations that are formed in
the water with ammonia and other substances that react with chlorine.
The fact that chlorine can be easily detected and measured makes chlorine a
favorite water sanitizer of those concerned with the public safety of water
supplies. Chlorine concentrations in the range of 0.1 to 0.4 parts per million
are usually maintained in municipal supplies.

SMART SPECTRO TEST PROCEDURES 05.04

Chlorine, Free–UDV 1/3