LaMotte SMART Spectro Spectrophotometer User Manual

Page 80

Advertising
background image

PROCEDURE–FREE CHLORINE

Use universal sample holder.

1.

Press and hold

ON button until spectrophotometer turns on.

2.

Scroll to and select PROGRAMMED TESTS.

3.

Scroll to and select ALL TESTS (or another sequence containing
15 Chlorine

) from TESTING MENU.

4.

Scroll to and select 15 Chlorine from menu.

5.

Rinse a clean tube (0290) with sample water. Fill to the 10 mL line with
sample.

6.

Insert tube into chamber, close lid and select SCAN BLANK.

7.

Remove tube from Spectro and pour off all but a sufficient amount of
sample water to cover a tablet. Add one *Chlorine DPD #1 Instrument
Grade Tablet (6903). Crush tablet with a tablet crusher (0175), then add
sample water until tube is filled to 10 mL line. Cap tube and shake until
tablet has dissolved. Solution will turn pink if free chlorine is present. Wait
15 seconds, but no longer than 30 seconds. Mix.

8.

Insert tube into chamber, close lid and select SCAN SAMPLE.

PROCEDURE–COMBINED CHLORINE

Use universal sample holder.

9.

Add one *Chlorine DPD #3 Instrument Grade Tablet (6197) to sample
from Step 8 above. Crush tablet with tablet crusher (0175). Cap tube and
shake until tablet dissolves. An increase in color represents combined
chlorine.
þNOTE: For wastewater samples, Standard Methods for the Examination of

Water and Wastewater recommends waiting 2 minutes for full color
development.

10.

Insert sample into chamber, close lid and select SCAN SAMPLE. Record
result as Total Chlorine.

11.

Subtract free chlorine reading from total chlorine reading to obtain
concentration of combined chlorine.

12.

Press the

OFF button to turn off the spectrophotometer or press the EXIT

button to exit to a previous menu or make another menu selection.

SMART SPECTRO TEST PROCEDURES 04.08

Chlorine–Bromine–Iodine 3/8