LaMotte SMART Spectro Spectrophotometer User Manual
Page 203
PROCEDURE
Use universal sample holder.
1.
Press and hold
ON button until spectrophotometer turns on.
2.
Scroll to and select PROGRAMMED TESTS.
3.
Scroll to and select ALL TESTS (or another sequence containing
79 Phosphate-H
) from TESTING MENU.
4.
Scroll to and select 79 Phosphate-H from menu.
5.
Rinse a clean tube (0290) with sample water. Fill to the 10 mL line with
sample.
6.
Insert tube into chamber, close lid and select SCAN BLANK.
7.
Remove tube from Spectro. Use the 1.0 mL pipet (0354) to add 2.0 mL of
*VM Phosphate Reagent (4410). Cap and mix. Wait 5 minutes for full
color development.
8.
After 5 minute waiting period, mix, insert tube into chamber, close lid and
select SCAN SAMPLE. Record result.
9.
Press
OFF button to turn spectrophotometer off or press EXIT button to
exit to a previous menu or make another menu selection.
Phosphate–HR 2/2
SMART SPECTRO TEST PROCEDURES 05.04