LaMotte SMART Spectro Spectrophotometer User Manual

Page 147

Advertising
background image

PROCEDURE

Use universal sample holder.

1.

Press and hold

ON button until spectrophotometer turns on.

2.

Scroll to and select PROGRAMMED TESTS.

3.

Scroll to and select ALL TESTS (or another sequence containing
46 H Peroxide-LR

) from TESTING MENU.

4.

Scroll to and select 46 H Peroxide-LR from menu.

5.

Rinse a clean tube (0290) with sample water. Fill to the 10 mL line with
sample.

6.

Insert tube into chamber, close lid and select SCAN BLANK. (See Note)

7.

Remove tube from Spectro and add 4 drops of *Hydrogen Peroxide
Reagent #1 (6452). Cap and mix.

8.

Add one *Hydrogen Peroxide LR Tablet (6454). Crush tablet with tablet
crusher (0175). Cap and mix for 30 seconds. Solution will turn pink if
hydrogen peroxide is present. Wait 5 minutes for full color development.

9.

At end of 5 minute waiting period, mix, insert tube into chamber, close lid
and select SCAN SAMPLE. Record result.

10.

Press

OFF button to turn spectrophotometer off or press EXIT button to

exit to a previous menu or make another menu selection.

þNOTES: For best possible results, a reagent blank should be determined to

account for any contribution to the test result by the reagent system. To
determine the reagent blank, follow the above test procedure to scan a
distilled or deionized water blank. Then follow the above procedure to
perform the test on a distilled or deionized water sample. This test result is
the reagent blank. Subtract the reagent blank from all subsequent test results
of unknown samples. It is necessary to determine the reagent blank only
when a new lot number of reagents are obtained.

For the most accurate results, the sample and reagents should be at 25±4°C.

Hydrogen Peroxide–LR 2/2

SMART SPECTRO TEST PROCEDURES 04.08

Advertising