LaMotte SMART Spectro Spectrophotometer User Manual

Page 123

Advertising
background image

PROCEDURE

Use universal sample holder.

1.

Press and hold

ON button until spectrophotometer turns on.

2.

Scroll to and select PROGRAMMED TEST from menu.

3.

Scroll to and select ALL TESTS (or another sequence containing
29 Copper BCA-LR

) from TESTING MENU.

4.

Scroll to and select 29 Copper BCA-LR from menu.

5.

Rinse a clean tube (0290) with sample water. Fill to the 10 mL line with
sample.

6.

Insert tube into chamber, close lid and select SCAN BLANK.

7.

Remove tube from Spectro and add one *Copper Tablet (T-3808). Cap
and shake vigorously until tablet dissolves. Solution will turn purple if
copper is present. Wait 2 minutes.

8.

At end of 2 minute waiting period, mix, insert tube into chamber, close lid
and select SCAN SAMPLE. Record result.

9.

Press

OFF button to turn spectrophotometer off or press EXIT button to

exit to a previous menu or make another menu selection.

Copper–LR 2/2