LaMotte SMART Spectro Spectrophotometer User Manual

Page 163

Advertising
background image

PROCEDURE

Use universal sample holder.

1.

Press and hold

ON button until spectrophotometer turns on.

2.

Scroll to and select PROGRAMMED TESTS.

3.

Scroll to and select ALL TESTS (or another sequence containing
56 Manganese H

) from TESTING MENU.

4.

Scroll to and select 56 Manganese H from menu.

5.

Rinse a tube (0290) with sample water. Fill to the 10 mL line with sample.

6.

Insert tube into chamber, close lid and select SCAN BLANK.

7.

Remove tube from Spectro. Use the 0.1 g spoon (0699) to add two
measures of Manganese Buffer Reagent (6310). Cap and mix until powder
dissolves.

8.

Use the 0.15 g spoon (0727) to add one measure of *Manganese Periodate
Reagent (6311). Cap and shake for one minute. An undissolved portion of
the reagent may remain in the bottom of the tube without adversely
affecting the test results. Wait two minutes for maximum color
development. Solution will turn pink if manganese is present.

9.

At the end of the two minute waiting period, mix, insert tube into
chamber, close lid and select SCAN SAMPLE. Record result.

10.

Press

OFF button to turn spectrophotometer off or press EXIT button to

exit to a previous menu or make another menu selection.

Manganese–HR 2/2

SMART SPECTRO TEST PROCEDURES 05.04

Advertising