The menu will indicate the spectro is blanking – LaMotte SMART Spectro Spectrophotometer User Manual

Page 38

Advertising
background image

n

ENTERING A NEW CALIBRATION

Scroll to NEW CALIBRATION. Press the

ENTER/Q button.

101 H20

NAME THE TEST

SELECT WAVELENGTH

* NEW CALIBRATION

FORMAT RESULT

The INPUT DATA NUMBERS menu appears.
Using the

s and t buttons set the number of

data points to be used. When finished press the

ENTER/Q button. The NEW CALIBRATION

menu appears.

INPUT DATA NUMBERS

DATA NUMBERS = 8

SET PRESS ENTER

The NEW CALIBRATION menu allows the entry
of the data points. Select ENTER STD.1 by
pressing

ENTER/Q button while the * is

adjacent to ENTER STD.1 in the display.

NEW CALIBRATION

* ENTER STD.1

ENTER STD.2

ENTER STD.3

ENTER STD.8

CALCULATE

Place a blank in the sample chamber. Press the

ENTER/Q button to SCAN BLANK. This is the

only choice.

ENTER STD.1

* SCAN BLANK

STD.1=

!0123456789.

The menu will indicate the Spectro is blanking.

ENTER STD.1

BLANKING

STD.1=

!123456789.

When BLANKING is finished the display will
indicate BLANKED and a blinking cursor will
appear on the character selection line.

ENTER STD.1

BLANKED

* STD.1=

!0123456789.

Continued on next page ...

38

SMART SPECTRO SPECTROPHOTOMETER 05.04