Phosphate–high range – LaMotte SMART Spectro Spectrophotometer User Manual

Page 202

Advertising
background image

PHOSPHATE–HIGH RANGE

VANADOMOLYBDOPHOSPHORIC ACID METHOD
CODE 3655-SC

QUANTITY

CONTENTS

CODE

2 x 60 mL

* VM Phosphate Reagent

*4410-H

1

Pipet, 1.0 mL, plastic

0354

*WARNING: Reagents marked with a * are considered to be potential health hazards. To view
or print a Material Safety Data Sheet (MSDS) for these reagents see MSDS CD or
www.lamotte.com. To obtain a printed copy, contact LaMotte by e-mail, phone or fax.

Phosphate treatments in boiler and cooling water and other industrial water
systems are run at levels up to 100 ppm orthophosphate. These high levels
permit the use of a simpler, high range test.

APPLICATION:

Boiler, cooling, and industrial waters.

RANGE:

0.0–70.0 ppm Phosphate

METHOD:

Orthophosphate reacts in acid conditions with ammonium
vanadomolybdate to form vanadomolybdophosphoric acid.
This yellow color is proportional to the concentration of
orthophosphate and is read colorimetrically.

SAMPLE
HANDLING &
PRESERVATION:

If the analysis cannot be performed the same day of
collection, the sample should be preserved by the addition of
2 mL of concentrated sulfuric acid or 40 mg mercuric
chloride per liter and refrigerated at 4°C.

INTERFERENCES:

Silica interferes only if the sample is heated. Arsenate,
fluoride, thorium, bismuth, sulfide, thiosulfate, and
thiocyanate cause negative interference.

SMART SPECTRO TEST PROCEDURES 05.04

Phosphate–HR 1/2

Advertising