PROCEDURE

Use universal sample holder

1.

Press and hold

ON button until spectrophotometer turns on.

2.

Scroll to and select PROGRAMMED TESTS.

3.

Scroll to and select ALL TESTS (or another sequence containing
21 Chloride-TT

) from TESTING MENU.

4.

Scroll to and select 21 Chloride-TT from menu.

5.

Rinse a tube (0290) with sample water. Fill to 10 mL with sample.

6.

Insert the tube into chamber, close lid and select SCAN BLANK.

7.

Remove the tube from Spectro.

8.

Add one *Chloride Spectrophotometric Grade Tablet (3885A).

9.

Use Tablet Crusher (0175) to crush tablet.

10.

Cap tube.

11.

Invert 2 times.

12.

Wait 3 minutes. Do NOT mix.

13.

Insert tube into chamber, close lid and select SCAN SAMPLE. Record
result in ppm chloride.

14.

Press

OFF button to turn the colorimeter off or press EXIT button to exit

to a previous menu or make another menu selection.

þNOTE: For best possible results, a reagent blank should be determined to

account for any contribution to the test result by the reagent system. To
determine the reagent blank, follow the above test procedure to scan a
distilled or deionized water blank. Then follow the above procedure to
perform the test on a distilled or deionized water sample. This test result is
the reagent blank. Subtract the reagent blank from all subsequent test results
of unknown samples.

The reagent system is temperature sensitive. The calibration is for 25

ºC. If

sample is at 30

ºC, multiply resulting ppm by 1.1. If the sample is at 20ºC,

multiply resulting ppm by 0.9.

Chloride - TesTab 2/2

SMART SPECTRO TEST PROCEDURES 04.08