PROCEDURE

Use universal sample holder

1.

Press and hold

ON button until spectrophotometer turns on.

2.

Scroll to and select PROGRAMMED TESTS.

3.

Scroll to and select ALL TESTS (or another sequence containing
66 Nitrate-TT

) from TESTING MENU.

4.

Scroll to and select 66 Nitrate-TT from menu.

5.

Rinse a tube (0290) with sample water. Fill to 10 mL with sample.

6.

Insert the tube into chamber, close lid and select SCAN BLANK.

7.

Remove the tube from Spectro.

8.

Add one *Nitrate Spectrophotometric Grade Tablet (3881).

9.

Use Tablet Crusher (0175) to crush tablet.

10.

Cap tube.

11.

Invert tube 60 times per minute for 2 minutes. (One inversion equals
180°).

12.

Wait 5 minutes. Do

NOT mix.

13.

Insert tube into chamber, close lid and select SCAN SAMPLE. Record
result in ppm nitrate.

14.

Press

OFF button to turn the spectrophotometer off or press EXIT button

to exit to a previous menu or make another menu selection.

þNOTES: For best possible results, a reagent blank should be determined to

account for any contribution to the test result by the reagent system. To
determine the reagent blank, follow the above test procedure to scan a
distilled or deionized water blank. Then follow the above procedure to
perform the test on a distilled or deionized water sample. This test result is
the reagent blank. Subtract the reagent blank from all subsequent test results
of unknown samples.

To convert nitrate (NO

3

) results to nitrate-nitrogen (NO

3

-N), divide by 4.4.

Nitrate-TesTab 1/2

SMART SPECTRO TEST PROCEDURES 04.08