Fig. 2.5. ShockPod with cuvette.
The plastic panel covering the electrode may be removed for cleaning should the electrode become
dirty. To clean the unit:
1. Disconnect the ShockPod from the Gene Pulser Xcell main unit.
2. Press the tab to open the chamber.
3. Using a phillips screwdriver, loosen the screw holding the safety panel (Figure 2.5). Lift the screw
and the safety panel out of the ShockPod by the lip on the front of the panel.
4. The area around the electrodes may be cleaned with a cotton swab and warm, soapy water. Dry
the electrodes using a dry cotton swab or Kimwipe.
5. Reassemble the unit by centering the safety panel on the base of the ShockPod and tightening the
screw.