Bio-Rad Gene Pulser Xcell™ Electroporation Systems User Manual

Page 53

Advertising
background image

Table 4.1 Droop associated with pulse length at various sample resistances for the
high-voltage and low-voltage ranges for the Gene Pulser Xcell.

High voltage circuit

Low voltage circuit

Droop %

10

20

5

10

20

Sample
Resistance

Pulse length (msec)

Pulse length (msec)

(ohms)

20

0.051 0.109 0.223

3.34

7.14

14.6

200

0.510 1.09 2.23

33.4 71.4 146

1000

2.55 5.45 11.2

167

357 730

3500

8.93 19.1 39.0

585 1249 2556

.

Fig. 4.1. (A) Exponential decay pulse from a capacitance discharge system. When a capacitor,
charged to an initial voltage V

o

, is discharged into cells, the voltage applied to the cells decreases

over time in an exponential surve such that the voltage “V” at any given time “t” is given by
V = V

o

e

–(t/RC)

. In the special case where t = CR then V

o

/e. The value CR is known as the time con-

stant of the voltage decay. The shorter the time constant the faster the decay. (B) Square wave
pulse from a capacitance discharge system. The pulse length is the time the cells are subjected
to the discharge. During the pulse the voltage again decreases in an exponential decay so that at
the end of the pulse the voltage is lower than at the beginning. We call this drop in voltage the

pulse droop and measure it as a percentage of the initial voltage.

47

Time

V

0

V

0

t

V

t

ττ

V

0

e

Time

V

olta

g

e

(V)

V

olta

g

e

(V)