6.
Set the spectrophotometer to 595 nm. Zero
the instrument with the blank sample (not
required for microplate readers). Measure the
absorbance of the standards and unknown
samples. Refer to Section 3 for data analysis.
Note: If the spectrophotometer has a
reference and sample holder, the instrument
can be zeroed with two blank samples. If the
effect of buffer on absorbance is required,
zero the instrument with a cuvette filled with
water and dye reagent in the reference
holder.
2.2 Microassay Protocol
1.
The microassay protocol can be performed in
two different formats, a 2 ml cuvette assay
14