Troubleshooting

•

Make sure that the room is well ventilated. CO

2

readings may be elevated in a closed

room.

•

Confirm that the calibration gas is of known concentration. If the CO

2

readings are

consistently elevated or depressed, the calibration gas may be suspect. Try another
gas source.

•

Make sure that the CO

2

module was properly set up and operating for the required

minimum period before testing.

•

Check for air leaks in the pneumatic tubing and connections. Repair or replace as
needed.

•

Confirm that the sample gas flow rate is greater than the CO

2

module’s pump flow

rate.

•

Confirm that silicon tubing is not used in the procedure (silicon absorbs CO

2

which

affects the gas concentration stability and accuracy of the procedure).