LC101 CO

2

Module OEM Implementation Manual

Section 4 - Host/Module Communications

Welch Allyn OEM Technologies

Confidential

Page 31

<Qxxy>

Function enable

Allows the host to enable or disable functions listed below.

<Qxxy>

where xx (ASCII) is the compensation and y (ASCII) is
ON/OFF status.

xx =

00h O

2

compensation

01h N

2

O compensation

02h desflurane compensation

08h water vapor compensation

09h BTPS compensation

0Ah baseline compensation

7Fh protect mode

y =

00 binary

ON

01 binary

OFF

3Fh

current ON/OFF status

Example
<Q011> compensate the CO

2

waveform data for N

2

O

<Q7F0> turn off protect mode (to allow certain commands)

<F20yy>

Change pump flow rate

Allows the host to specify the pump flow rate in milliliters per
minute.

<F20yy> where yy (8 bit) is the current pump flow rate.

yy range

90 – 200 ml/min (5Ah – C8h)

yy default

175 ml/min (AFh)

Example
<F205A> set the pump flow rate to 90 ml/minute

Note: Using this command to set a desired flow rate, the pump flow
rate will revert back to default settings upon power up. The flow rate
can be fixed in the EEPROM to a specific default setting. See
EEPROM Map for location details.

<C80>

Request software reset

Allows the host to reset the Module. This command is only
accepted if the Module is in the unprotected mode. See Function
enable <Qxxy>.