Page 45

Host Command/Module Response Overview Table

Command

Command Description

Response

Response Description

<M21>

Enter standby mode

<Sxx06>

Acknowledge with Module status

<M23>

Enter measurement mode

<Sxx06>

Acknowledge with Module status

<M24>

Enter auto run mode

<Sxx06>

Acknowledge with Module status

(Invoked with enter measurement
or auto run mode command)

<Wxxyy>

CO

2

waveform data packets sent at

programmed interval

(Invoked with enter measurement
or auto run mode command)

<Zxxyyww>

Breath data packets sent at pro-
grammed interval

<C00>

Request Module status

<Sxxyy>

Acknowledge with Module status

<C10>

Sample Zero Reference Gas

<Sxx25>

Status 25 = Cal in Progress

<C11>

Sample Span Reference Gas

<Sxx25>

Status 25 = Cal in Progress

<C12>

Perform Zero Calibration

<Sxx25>

Status 25 = Cal in Progress

<C13>

Abort Calibration

<Sxx00>

<Sxx22>

Status 00, ok (operation aborted)

Calibration not in Progress

<C20>

Request software version

<Vxxxmmd-
dyyyy>

Version # with month, day, year sent

<C21>

Request hardware version

<Hxx>

Hardware version # sent

<C22>

Request ambient barometric
pressure

<Lxxxx>

Ambient barometric pressure sent

<C23>

Request sensor temperature

<Txx>

Sensor temperature sent

<C26>

Request single CO

2

packet

<wxxyy>

CO

2

measurement sent

<C27>

Request single breath data
packet

<zxxyyww>

Breath data packet sent

<C2E>

Request current pump flow rate

<Bxxxx>

Current pump flow rate sent

<C31>

Request sensor EEPROM revi-
sion #

<Rxx>

Revision # sent

<C32>

Request customer code

<Ixx>

Sensor customer code sent

<C33>

Request sensor serial #

<Nxxxx>

Sensor serial # sent

<C3A>

Request last calibration date

<Dmmd-
dyyyy>

Last calibration date sent

<C80>

Software reset

No response

<Axx>

Specify breath data timeout

<axx>

Data field echoed

<F20yy>

Specify pump flow rate

<f20yy>

Command echoed

<Gxx>

Update Span Gas

<Sxx24>

Status 24 = Cal Ready for Next Step

<Lmmd-
dyyyy>

Update Calibration Date.

<Sxx24>

Status 24 = Cal Ready for Next Step

<Nxx>

Specify CO

2

waveform data

packet interval

<nxx>

Data field echoed

<Oxx>

Specify breath data packet inter-
val

<oxx>

Data field echoed

<Qxxy>

Specify measurement compen-
sation ON or OFF, SW reset

<qxxy>

New and previous measurement
compensation settings echoed