Welch Allyn LC101 CO2 Module - User Manual User Manual

Page 37

Advertising
background image

LC101 CO

2

Module OEM Implementation Manual

Section 4 - Host/Module Communications

Welch Allyn OEM Technologies

Confidential

Page 37

<C27>

Request single breath packet

<zxxyyww>

xx = ETCO

2

mmHg

yy = RR bpm

ww = InsCO

2

mmHg

Example:
<z201500>

ETCO

2

= 32

RR= 21

InsCO2= 0

<C2E>

Request current pump flow rate

<Bxxxx>

xxxx = pump flow rate ml/min

Example:
<B0096>

flow rate = 150 ml/min

<C31>

Request sensor EEPROM revi-
sion #

<Rxx>

xx = revision #

Example:
<R02>

revision # = 2

<C32>

Request sensor customer code

<Ixx>

xx = customer code

Example:
<I00>

customer code = 00

<C33>

Request sensor serial #

<Nxxxx>

xxxx = serial #

Example:
<N34C2>

serial # = 13506

<C3A>

Request last calibration date

<Dmmddyyyy>

mm = month

dd = day

yyyy = year

Example:
<D06041998>

last cal date= 6/4/1998

Advertising