Module OEM Implementation Manual

Page 70

Confidential

Welch Allyn OEM Technologies

29, Unable to Perform Calibration, Data Error

The acquired CO

2

data is invalid or out of range for the calibration. The Module will abort the

calibration. This error also occurs when calibration is aborted due to any other non-fatal error.

2A, Bad Calibration CRC

The system EEPROM CRC is not valid after being updated with the new constants.

Sensor Faults

CO

2

sensor faults are conditions that result in shutting down the sensor. A CO

2

sensor shutdown

turns off the lamp and the pump.

To clear a sensor fault, perform a software reset followed by a new host command.

If the Module

does not accept the command, a hard reset is required.

Sensor faults are listed in the following table.

51, CO

2

Sensor Manufacturer Code Mismatch

Sensor manufacturer code stored in sensor EEPROM does not match the host monitor
manufacturer code.

•

Verify the data in the EEPROM location.

Status
Message Code
(hex)

Definition

Possible Causes

51

Customer code mismatch

• Incorrect EEPROM data or bad sensor EEPROM

57

CO

2

sensor not found

• Main Board not acknowledging the sensor

• loose or faulty sensor connection

60

CO

2

sensor EEPROM error

in revision #

• incompatible/invalid sensor EEPROM revision #

• faulty sensor EEPROM

65

CO

2

sensor EEPROM error

in read or write operation

• invalid read/write to sensor EEPROM;

• faulty sensor EEPROM

66

CO

2

sensor EEPROM error

in CCITT/CRC code

• invalid sensor EEPROM CCITT/CRC code;

• faulty sensor EEPROM

70

CO

2

sensor temperature

too high

• sensor exposure to excessive heat source;

• faulty sensor

71

CO

2

sensor temperature

too low

• sensor exposure to excessive cold source;

• faulty sensor