LC101 CO

2

Module OEM Implementation Manual

Page 74

Confidential

Welch Allyn OEM Technologies

•

Check the exhaust tubing for an improper connection to a vacuum source.

•

Refer to Section 5, Pneumatic Operation, for additional information.

84, Barometric Pressure Too High

Barometric pressure transducer reading is greater than the upper barometric limit.

•

Check that the host monitor is operating at barometric pressures less than 824 mmHg
(1,300 ft below sea level).

•

Possible transducer error.

85, Barometric Pressure Too Low

Barometric pressure transducer reading is less than the lower barometric limit.

•

Check that the host monitor is operating at barometric pressures greater than 408
mmHg (15,500 ft above sea level).

•

Possible transducer error.