Calibration command responses – Welch Allyn LC101 CO2 Module - User Manual User Manual

Page 40

Advertising
background image

LC101 CO

2

Module OEM Implementation Manual

Page 40

Confidential

Welch Allyn OEM Technologies

Calibration Command Responses

Refer to the Status Message Code Table and the Host Command/Module Response Overview
Table at the end of this section for message descriptions.

Host Command

Module Response

Module Response Format

<Lmmddyyyy>

Update calibration date

<Sxxyy>

xx = mode

yy = message
Example:
<S6424>

64= Autorun mode

24= Cal. ready for next step
Example:
<S6421>

64= Autorun mode

21= Cal. already in progress

<Gxx>

Update span gas concen-
tration

<Sxxyy>

xx = mode

yy = message
Example:
<S6424>

64= Autorun mode

24= Cal. ready for next step
Example:
<S6422>

64= Autorun mode

22= Cal. not in progress

<C10>

Sample zero reference gas

<Sxxyy>

xx = mode

yy = message
Example:
<S6425>

64= Autorun mode

25= Cal. in progress

<C11>

Sample span gas

<Sxxyy>

xx = mode

yy = message
Example:
<S6425>

64= Autorun mode

25= Cal. in progress

<C13>

Abort calibration

<Sxxyy>

xx = mode

yy = message
Example:
<S6400>

64= Autorun mode

00= OK (operation aborted)