Module OEM Implementation Manual

Section 4 - Host/Module Communications

Welch Allyn OEM Technologies

Confidential

Page 35

Status Response Message Byte

<Sxxyy>

where yy

=

Status message

(2 Hex digits = 2 ASCII characters)

Status Response message byte (yy) contains:

generic status messages (00h – 06h)

fault status messages (10h – 85h)

Refer to the Status Message Table at the end of this section for message descriptions. More
detailed information about fault messages is also available in Appendix C, Error Messages &
Recovery.

Mode Command Responses

Host Command

Module Response

Module Response Format

<M21>

Enter standby mode

<S6106>

xx = mode

yy = message
Example:
<S6106>

61= Standby mode

06= Acknowledge mode command

<M23>

Enter measurement mode

<S6306>

xx = mode

yy = message
Example:
<S6306>

63= Measurement mode

06= Acknowledge mode command

<M24>

Enter auto run mode

<S6406>

xx = mode

yy = message
Example:
<S6406>

64= Autorun mode

06= Acknowledge mode command