Welch Allyn LC101 CO2 Module - User Manual User Manual
Page 39

LC101 CO
2
Module OEM Implementation Manual
Section 4 - Host/Module Communications
Welch Allyn OEM Technologies
Confidential
Page 39
<Zxxyyzz>, Breath Data
Includes 6 bytes of breath data that are sent at the programmed interval during measurement
and auto run modes.
<Zxxyyzz> where xxyyzz:
xx
ETCO
2
mmHg
2*(ASCII)
yy
RR bpm
2*(ASCII)
zz
InsCO
2
mmHg
2*(ASCII)
Range:
ETCO
2
0 – 99 mmHg (00h – 63h)
RR
0 – 250 bpm (00h – FAh)
InsCO
2
0 – 99 mmHg (00h – 63h)
Example:
ETCO
2
= 39 mmHg / RR = 12 bpm / InsCO
2
= 0 mmHg
Additional Configuration Command Responses
{STX}
“Z”
xx
yy
zz
CCITT/CRC
{ETX}
breath byte string
0x27
0x0C
0x00
0x1A
{STX}Z270C001A{ETX}
02h
5Ah
32h37h
30h43h
30h30h
31h41h
03h
Host Command
Module Response
Description
<Axx>
Change breath data clear rate
<axx>
response = echo data field
<Nxx>
Change waveform update rate
<nxx>
response = echo data field
<Oxx>
Change breath data update rate
<oxx>
response = echo data field
<Qxxy>
Function enable
<qxxy>
response = echo new/current status
<F20yy>
Change pump flow rate
<F20yy>
response = echo data field
<C80>
Request software reset
no response