Welch Allyn LC101 CO2 Module - User Manual User Manual

Page 39

Advertising
background image

LC101 CO

2

Module OEM Implementation Manual

Section 4 - Host/Module Communications

Welch Allyn OEM Technologies

Confidential

Page 39

<Zxxyyzz>, Breath Data

Includes 6 bytes of breath data that are sent at the programmed interval during measurement
and auto run modes.

<Zxxyyzz> where xxyyzz:

xx

ETCO

2

mmHg

2*(ASCII)

yy

RR bpm

2*(ASCII)

zz

InsCO

2

mmHg

2*(ASCII)

Range:

ETCO

2

0 – 99 mmHg (00h – 63h)

RR

0 – 250 bpm (00h – FAh)

InsCO

2

0 – 99 mmHg (00h – 63h)

Example:

ETCO

2

= 39 mmHg / RR = 12 bpm / InsCO

2

= 0 mmHg

Additional Configuration Command Responses

{STX}

“Z”

xx

yy

zz

CCITT/CRC

{ETX}

breath byte string

0x27

0x0C

0x00

0x1A

{STX}Z270C001A{ETX}

02h

5Ah

32h37h

30h43h

30h30h

31h41h

03h

Host Command

Module Response

Description

<Axx>

Change breath data clear rate

<axx>

response = echo data field

<Nxx>

Change waveform update rate

<nxx>

response = echo data field

<Oxx>

Change breath data update rate

<oxx>

response = echo data field

<Qxxy>

Function enable

<qxxy>

response = echo new/current status

<F20yy>

Change pump flow rate

<F20yy>

response = echo data field

<C80>

Request software reset

no response

Advertising