Appendix e specifications, Lc101 main board, Appendix e – Welch Allyn LC101 CO2 Module - User Manual User Manual
Page 77: Specifications

LC101 CO
2
Module OEM Implementation Manual
Welch Allyn OEM Technologies
Confidential
Page 77
Appendix E
Specifications
LC101 Main Board
PHYSICAL SPECIFICATIONS
Size
3.15 in. x 3.90 in. (80.3 mm x 99.3 mm)
Clearance
0.85 in. (22 mm) above board, 0.12 in (3 mm) below board
Orientation
No limitations
Weight
Less than 1 pound (454 grams) with full pneumatics.
SYSTEM PERFORMANCE SPECIFICATIONS
Sampling Method
Built-in sidestream sensor
Patient Application
Neonatal to adult
CO
2
Concentration Display Range
0 - 99 mmHg
Respiration Rate Range
1 - 99 breaths per minute
Typical CO
2
Accuracy
a
Start-Up Time
Less than 10 seconds to acquire CO
2
waveform data
Less than 3 minutes to full operating specification.
Calibration
Zero Calibration and Two-Point User Calibration
C-Cap Rise Time
Less than 200 msec (10% - 90%, bench only)
Flow Rate Range
User-selectable, variable from 90 to 200 ml/min (defaults to 175
ml/min)
Flow Rate Accuracy
-20/+15% of set value.