Watertrap Receiver Switch
The watertrap receiver switch is wired in the “normally open” configuration.
The watertrap receiver switch connects directly to the Main Board via a wiring harness and
connector. The watertrap receiver switch assembly is equipped with a 2 pin connector (Molex
P/N 22-01-3027) which electrically connects directly to the Main Board at J102.
The watertrap receiver switch wiring meets typical insulation standards but does not meet patient
isolation requirements. When applicable, harness routing must not violate patient isolation
requirements. Refer to Section 6, Regulatory, for additional regulatory and safety information.
Exhaust Tubing
In the event of an exhaust tubing malfunction, gas buildup can occur within the host system.
WARNING To protect against potentially flammable gas buildup, the tubing must be
single-fault protected. Refer to Section 6, Regulatory, for additional regulatory
and safety information.