40, Watchdog Error

Watchdog timeout has occurred.

44, System EEPROM Error/CCITT/CRC Code

CCITT/CRC code read from the system EEPROM is invalid.

46, System FLASH ROM Error/CCITT/CRC Code

CCITT/CRC code read from the system FLASH ROM is invalid.

47, System Communication Error

4B, System External RAM Error

4C, System Internal RAM Error

4D, System FLASH ROM Error/Checksum

Checksum read from the system FLASH ROM is invalid.

4E, Stack Overflow Error

An error in stack usage occurred.

4F, System SW Error/Main Program

Main SW program is exited.

80, Pump Fault

High flow and high vacuum conditions are detected in the pneumatics.

Refer to Section 5, Pneumatic Operation, for information about this fault.

81, Unexpected Reverse Flow

An unexpected reverse flow condition in the pneumatics, indicated by non-zero initial values,
turns the pump off.

Check the watertrap and cannula for an improper connection to a vacuum source.

Check the exhaust tubing for an improper connection to a positive pressure source.

Refer to Section 5, Pneumatic Operation, for additional information.

82, Unexpected Forward Flow

An unexpected forward flow condition in the pneumatics, indicated by non-zero initial values,
turns the pump off.

Check the watertrap and cannula for an improper connection to a positive pressure
source.