The Configuration menu provides a means to setup up how the CO
2
Monitor functions. The
configuration items are saved and restored at start up of the CO
2
Monitor application. The
Configuration menu contains the following items:
Configure\ Mode\ Mainstream (Duet module only) - Checking this enables mainstream
operation and disables sidestream operation.
Configure\ Mode\ Sidestream (Duet module only) - Checking this enables sidestream
operation and disables mainstream operation.
Configure\ Display\Units - The CO
2
Monitor will display CO
2
values in either partial pressure
(mmHg) or percent. These units apply to the waveform and the EtCO
2
and InsCO
2
values.
Configure\ Display\Sweep Rate - The sweep rate setting determines the sweep rate of the CO
2
waveform. This actual sweep rate may vary depending on the PC’s configured display resolution.
Configure
Mode
Mainstream
Sidestream
Display
CO2 Module
Units
Sweep Rates
CO2 Scale
Gridlines
Waveform Fill
mmHg
Percent
6.25 mm/sec
12.5 mm/sec
25 mm/sec
50 mmHg
75 mmHg
100 mmHg
14%
10%
6%
Flow Rate
CommPort
90 ml/min
175 ml/min
150 ml/min