Zero calibration sequence example – Welch Allyn LC101 CO2 Module - User Manual User Manual
Page 57

LC101 CO
2
Module OEM Implementation Manual
Appendix A - Software Prodedures
Welch Allyn OEM Technologies
Confidential
Page 57
1.
Initialize breath data update rate <O00>.
2.
Request ambient barometric pressure <C22>.
3.
Divide ETCO
2
(in mmHg) value by ambient barometric pressure (in mmHg). The result is the
ETCO
2
value as a percentage of the barometric pressure.
Zero Calibration Sequence Example
Seq.
Host
Module
Description
1
<Q7F0>
Host sends Unprotect command.
<q7F0>
Module responds to command from host.
2
<C12>
Host sends
Perform Zero Calibration command.
<S6325>
<S6X03>
<S6321>
<S6323>
Module responds with status -
Calibration in Progress, or
if not in unprotected measurement or autorun mode, response is
Invalid
Operation (use <Q7F0>), or
if the previous calibration is active, response is
Cal. Already in Progress,
or
if the LC101 had been operating for < 5 min., response is
Cal. Low Run
Time.
3
<C00>
Host polls for completion of sampling.
<S6325>
<S6326>
<S6329>
Module responds with status –
Calibration in Progress. Repeat this
step until a response other than <S6325> is returned.
Module responds with status –
Calibration OK, or
if acquiring data does not meet specifications, response is
Unable to
Perform Cal., Data Error.
4
<Q7F1>
Host sends Protect command.
<q7F1>
Module responds to command from host.