Zero calibration sequence example – Welch Allyn LC101 CO2 Module - User Manual User Manual

Page 57

Advertising
background image

LC101 CO

2

Module OEM Implementation Manual

Appendix A - Software Prodedures

Welch Allyn OEM Technologies

Confidential

Page 57

1.

Initialize breath data update rate <O00>.

2.

Request ambient barometric pressure <C22>.

3.

Divide ETCO

2

(in mmHg) value by ambient barometric pressure (in mmHg). The result is the

ETCO

2

value as a percentage of the barometric pressure.

Zero Calibration Sequence Example

Seq.

Host

Module

Description

1

<Q7F0>

Host sends Unprotect command.

<q7F0>

Module responds to command from host.

2

<C12>

Host sends

Perform Zero Calibration command.

<S6325>

<S6X03>

<S6321>

<S6323>

Module responds with status -

Calibration in Progress, or

if not in unprotected measurement or autorun mode, response is

Invalid

Operation (use <Q7F0>), or
if the previous calibration is active, response is

Cal. Already in Progress,

or

if the LC101 had been operating for < 5 min., response is

Cal. Low Run

Time.

3

<C00>

Host polls for completion of sampling.

<S6325>

<S6326>

<S6329>

Module responds with status –

Calibration in Progress. Repeat this

step until a response other than <S6325> is returned.

Module responds with status –

Calibration OK, or

if acquiring data does not meet specifications, response is

Unable to

Perform Cal., Data Error.

4

<Q7F1>

Host sends Protect command.

<q7F1>

Module responds to command from host.