LC101 CO

2

Module OEM Implementation Manual

Page 36

Confidential

Welch Allyn OEM Technologies

Simple Command Responses

Host Command

Module Response

Module Response Format

<C00>

Request Module status

<Sxxyy>

xx = mode

yy = message
Example:
<S6400>

64= Autorun mode

00= OK status

<C20>

Request software version

<Vxxxmmddyyyy>

xxx = version #

mm = version month

dd = version day

yyyy = version year
Example:
<V13010231998>

version = 1.30

date = 10-23-1998

<C21>

Request hardware version

<Hxx>

xx = hardware version
Example:
<H25>

hardware version = 2.5

<C22>

Request stored ambient baro-
metric pressure reading from
absolute pressure transducer

<Lxxxx>

xxxx = amb baro pressure mmHg
Examples:
<L02E9>

amb baro pressure = 745 mmHg

<L02F8>

amb baro pressure = 760 mmHg

<C23>

Request sensor temperature

<Txx>

xx = (temperature x 4)

o

C

Example:
<T2A>

temperature = (42/4) = 10.5

o

C

<C26>

Request single CO

2

packet

<wxxyy>

xxyy = ppCO

2

x 256

CO

2

= 16 bit binary integer

Example:
<w1C5B>

waveform = 28.35

Advertising