Simple Command Responses
Host Command
Module Response
Module Response Format
<C00>
Request Module status
<Sxxyy>
xx = mode
yy = message
Example:
<S6400>
64= Autorun mode
00= OK status
<C20>
Request software version
<Vxxxmmddyyyy>
xxx = version #
mm = version month
dd = version day
yyyy = version year
Example:
<V13010231998>
version = 1.30
date = 10-23-1998
<C21>
Request hardware version
<Hxx>
xx = hardware version
Example:
<H25>
hardware version = 2.5
<C22>
Request stored ambient baro-
metric pressure reading from
absolute pressure transducer
<Lxxxx>
xxxx = amb baro pressure mmHg
Examples:
<L02E9>
amb baro pressure = 745 mmHg
<L02F8>
amb baro pressure = 760 mmHg
<C23>
Request sensor temperature
<Txx>
xx = (temperature x 4)
o
C
Example:
<T2A>
temperature = (42/4) = 10.5
o
C
<C26>
Request single CO
2
packet
<wxxyy>
xxyy = ppCO
2
x 256
CO
2
= 16 bit binary integer
Example:
<w1C5B>
waveform = 28.35