Page 71

57, CO

2

Sensor Not Found

CO

2

Bench, or C-Cap, not present on Main Board.

60, CO

2

Sensor EEPROM Error/Revision #

Revision # read from the CO

2

sensor EEPROM is not in the 1.xx format.

65, CO

2

Sensor EEPROM Error/Read/Write

A read or write error to the CO

2

sensor EEPROM has occurred.

66, CO

2

Sensor EEPROM Error/CCITT/CRC Code

CCITT/CRC code read from the CO

2

sensor EEPROM is invalid.

70, CO

2

Sensor Temperature Too High

The CO

2

sensor’s temperature is greater than 60

o

C for more than 30 seconds.

•

Check for the presence of an external excessive heat source that is in close proximity
to the Module. Remove the source or relocate the Module.

71, CO

2

Sensor Temperature Too Low

The CO

2

sensor’s temperature is less than 5

o

C for more than 30 seconds.

•

Check for excessive cold ambient conditions or the presence of an external cold
source that is in close proximity to the Module. Remove the source or relocate the
Module.