Data Processing and Operation Modes

TIMER mode

Even without the DRQ input from an external device, the sensor amplifier can periodically output
BCD data.
The DRQ input is ignored.

Time chart

(1) An external device inputs "ID No. 01" as the ID No. selection.
(2) The external device imports BCD output from ID No. 01 with an update cycle that starts no more

than 2 ms later.

(3) At the next BCD update cycle, if the ID No. has not changed, the BCD output for the same ID No.

is updated.

(4) An external device inputs "ID No. 02" as the ID No. selection.
(5) The external device imports BCD output from ID No. 02 with an update cycle that starts no more

than 2 ms later.

ID No. selection
input

(1) ID No. 01 selection

(4) ID No. 02 selection

BCD output

2 ms.
max.

(2), (3) ID No. 01 BCD data

(5) ID No. 02 BCD data

Strobe output

OFF

OFF

OFF

OFF

ON

ON

ON

ON

T/2

T

T

T

T/2

T/2

2 ms.
max.

2 ms min.

2 ms min.

T :BCD output update cycle (Set T with the advanced setting switches Nos. 4 to 6. For information on
advanced setting switches, refer to "(1) Advanced setting switches" (Page 7).)