4-3 Calibration Function

4-45

IB-E

4

Set
ting V

a

rious

F

unct

ions

Setting Method (Logical Correction)

Point

Logical correction cannot be set if “dimension mode” is selected for
“1. Measurement Mode”.

Main screen

1

Press the [MODE] and buttons for approx. 2
seconds on the main screen at the same time.

[30.CLb] is displayed on the main display (upper level).

Press the [MODE]

+  buttons for 2

seconds or more

at the same time

2

Press the / button to set [LoGic] and press the
button.

3

Enter “R.V. before calibration (1)” on the main display
using the
/ buttons. The “R.V. before calibration
(1)” blinks on the main display (upper level). [
SET1] is
displayed on the sub display's screen (lower level).

4

Press the [SET] button.

[set1] and the “R.V. you wish to display (1)” are alternately
displayed on the sub display's screen (lower level).

[SET] button

set2

5

Enter “R.V. you wish to display (1)” on the main
display using the
/ buttons.

6

Press the [SET] button. (SET1 confirmation operation)

After [set1] blinks several times on the sub display's screen
(lower level), [set2] is displayed.

[SET] button



LASER

BANK

0

1

2

3

HI

GO

LO

R.V. ANALOG

HI SHIFT

ZERO SHIFT

TIMING

LO

ALIGNMENT



HOLD CALC CHECK

F'(.6

LASER

BANK

0

1

2

3

HI

LO

R.V.

ANALOG

HI

SHIFT

ZERO SHIFT

TIMING

LO

ALIGNMENT

%.D

GO

HOLD

CALC

CHECK

Calibration setting

Setting value

Description

Default value

defLt

No correction (Default setting)

MEAS

Measured correction

LoGic

Logical correction

5'6

LASER

BANK

0

1

2

3

HI

GO

LO

R.V. ANALOG

HI SHIFT

ZERO SHIFT

TIMING

LO

ALIGNMENT



HOLD CALC CHECK

R.V. before calibration (1)

" 1. Measurement Mode" (page 4-8)

Item

Setting range

Default value

R.V. before
calibration (1)

- 999.99 to 999.99

0.00



LASER

BANK

0

1

2

3

HI

GO

LO

R.V. ANALOG

HI SHIFT

ZERO SHIFT

TIMING

LO

ALIGNMENT



HOLD CALC CHECK

R.V. you wish to display (1)

" 1. Measurement Mode" (page 4-8)

Item

Setting range

Default value

R.V. you wish to
display (1)

- 999.99 to 999.99

0.00