] filling performance test – Flintec FT-13 User Manual

Page 32

Advertising
background image

FT-13 Technical Manual, Rev. 1.40 March 2013

Page 32 of 56

[508 ] Maximum filling time

[ XXXXXX ]

This parameter controls the filling / discharging process time. If the filling / discharging
time exceeds this time period, an error message appears [FıLL t]. If it is set to “0”,
there will be no control. It can be entered as a multiple of 100 ms.
Factory default setting is 0.

[509 ] Maximum weight

[ XXXXXX ]

This parameter defines the maximum allowed weight value in gross. If the displayed
weight or the sum of the target weight and the tare weight is higher than the defined
value the message [OVER] is shown on the display. Factory default setting is 3000 kg.

[50A X ] Enable <Start> / <Reset> keys

0 : Disabled (factory default setting)

1 : Enabled

[50b X ] Filling quantity control

0 :

The filling quantity entered by the <Q> key decreases automatically after each filling cycle and
a warning message appears when it is zero (factory default setting).

1 :

The filling quantity is increased by one after each filling cycle and there will be no filling quantity
control.

[50C X ] Feeding type

For 2-speed feeding:

0 :

Coarse and fine feedings are enabled sequentially

1 :

Coarse & fine feedings and fine feeding are enabled sequentially

For 3-speed feeding:

2 :

Coarse, middle and fine feedings are enabled sequentially

3 :

Coarse & fine feedings, middle & fine feedings and fine feeding are enabled sequentially

4 :

Coarse & middle & fine feedings, middle & fine feedings and fine feeding are enabled
sequentially

[50d XX] Fine feeding time at cycle start

This parameter defines the fine feeding time before the coarse feeding. If this parameter is set to “0”, the
filling operation starts with the coarse feeding. It can be defined up to 99 s.

[50E ] Zeroing period

[ XXXXXX ]

This parameter defines the automatic zeroing period in [s]. The zeroing is performed
automatically at the next filling cycle after this time period.
For legal for trade applications the value entered here shall be less than the value
calculated at the OIML R-61 section A.5.3.5.
If the zeroing will be done each cycle, this value shall be entered less than the
minimum filling time. The allowed value range goes up to 5399 s (= 90 minutes).

[50F X.X ] Zeroing delay

This parameter defines the zeroing delay time before zeroing. It can be defined up to 9.9 s. The factory
default is 3.0 s.

[51-]

Increasing or decreasing indication

0 : Increasing (factory default setting)

1 : Decreasing

[59-] Filling Performance Test

This test is very helpful to learn the material flow and to define coarse and fine feeding cut-off values.

[590 ] Coarse feeding test

Press the <Enter> key to go into this menu. The [StArt ] message appears on the display.
The coarse feed starts by the <Start> key hold pressed for one second. After test feeding the display
shows the quantity of the material coarse lead in as [ XXXXXX ].

[591 ] Fine feeding test

Press the <Enter> key to go into this menu. The [StArt ] message appears on the display.
The fine feed starts by the <Start> key hold pressed for one second. After test feeding the display shows
the quantity of material fine lead in as [ XXXXXX ].