Profibus option, Electrical connections, Setup – Flintec FT-13 User Manual

Page 40: Data structure, 4 profibus option, 1 electrical connections, 2 setup [0--] interface block [03-] profibus, 3 data structure, 032 x ] data format, 033 xx] address

Advertising
background image

FT-13 Technical Manual, Rev. 1.40 March 2013

Page 40 of 56

8.4 Profibus Option

Indicators which are equipped with the Profibus option can be connected to a PLC bus as described below.

Profibus DP V0 & V1

Automatic (96 kbit/s to 12 Mbit/s)

Update rate

Weight data will be updated each 40 ms

Response time on entries

5 ms to 25 ms

8.4.1 Electrical Connections

The pin configuration of the female 9-pol D-Sub Profibus connector (DB9F) is described in the table below:

1

Operation mode LED

2

Status LED

3

Profibus Connector

Pin no.

Signal

Description

1

2

3

B-Line

Positive RxD / TxD, RS485 level

4

RTS

Request to send

5

GND

Isolated GND

6

+5 V DC

Isolated +5V output for bus termination

7

8

A-Line

Negative RxD / TxD, RS485 level

9

Housing

Shield

Bus cable shield, connected to Ground

Operation mode LED

State

Indication

Action / Comment

Off

Not on-line /No power

Check power and cable

Green

On-line, data exchange

Flashing Green

On-line, clear

Flashing Red (1 flash)

Setup error

Not supported

Flashing Red (2 flash)

PROFIBUS configuration error

Check GSD file

Status LED

State

Meaning

Off

No power or Profibus module is initialising

Green

Initialized

Flashing Green

Initialized and diagnostic event is present

Red

There is an exception error

8.5.2 Setup

[0--] Interface Block

[03-] Profibus

This sub-block includes the parameters of the Profibus port.

[032 X ] Data Format

0 : Signed 32 bit integer

1 : Signed 32 bit floating point

[033 XX] Address

The address of the instrument will be entered via numerical keypad. Value range is 1 to 126. After editing
the address it is required to restart the instrument.

The GDS file is available on a CD which is supplied together with the instrument.

After the setup of these parameters and the hardware connection as described above, the data interface can
be used as described in Appendix 4.

Important hint: During the setup the Profibus output (weight value and status bits) will not be updated.

8.5.3 Data Structure

See Appendix 4.