Appendix 3: modbus rtu data structure, Ppendix, Odbus – Flintec FT-13 User Manual

Page 46: Tructure

Advertising
background image

FT-13 Technical Manual, Rev. 1.40 March 2013

Page 46 of 56

A

PPENDIX

3:

M

ODBUS

RTU

D

ATA

S

TRUCTURE

If the instrument is programmed for Modbus, it can be used as a Modbus RTU slave on a RS485
communication line or on the Ethernet communication line. The Modbus slave adress is defined in the
parameter [Address]. Function code “0x03” and function code “0x10” are supported.

Read

command

Modbus

register

Function

code

Register

address

No. of

registers

Definition

Display value

40 001

03

00,00

00,02

Current display value; Gross or Net

Status

40 003

03

00,02

00,01

Bit0

Busy

1

– System Busy

Bit1

Data ok

1

– Data ok // 0 – Error

Bit2

Motion

1

– Weight is not stable

Bit3

Net Mode

1

– Net mode

Bit4…12

0

Bit13…15

Transfer
binary
(Bit13 to
Bit15) to
decimal and
compare to
error code

0

No Errors

1

ADC out of range

2

ADC overrange

3

ADC underrange

4

System fault

5

FT not in weighing or filling
mode

Tare weight

40 004

03

00,03

00,02

Current tare weight

Gross Weight

40 006

03

00,05

00,02

Current gross weight

Operation Mode

40 010

03

00,09

00,01

1 = Filling mode, 0 = Weighing mode

Target

40 011

03

00,0A

00,02

Current target weight

Tolerance

40 013

03

00,0C

00,02

Current tolerance weight

Coarse

40 015

03

00,0E

00,02

Current coarse value

Fine

40 017

03

00,10

00,02

Current fine value

CN

40 019

03

00,12

00,01

Current consecutive Number

Last filling value

40 020

03

00,13

00,02

Last filling weight

Quantity

40 022

03

00,15

00,01

Current quantity

Total

40 023

03

00,16

00,02

Current total weight

Inputs

40 025

03

00,18

00,01

Current input status

Outputs

40 026

03

00,19

00,02

Current output status:
Active output no. = Data bit no. se

t to “1”

(e.g. only Bit 1 + Bit 8 = “1”  only output 1 + 8 active)

Middle

40 028

03

00,1A

00,02

Current middle value

Calibration

Status

40 033

03

00,1F

00,01

D0…D7
Calibration
Process
Status

1

Ready for calibration

3

Zero calibration in process

4

Span calibration in process

9

Error (Refer to D8 ... D15 )

D8…D15
Calibration
Errors

1

Calibration Timeout

Restart calibration

2

ADC Error
– Re-energize the instrument
– If seen again, change board.

3

Instrument cannot be calibrated

Check load cell cable
– Re-energize the instrument

34

Instrument cannot be calibrated

– Load

cell signal is too low or too high

35

Calibration Error
– Calibration test weight is too small,
increase calibration weight value
(40031-32)
– Check load cell connections

37

Scale unstable
– Wait until scale becomes stable
– Check grounding wiring

38

Calibration Jumper is not installed-
Check Calibration jumper







Advertising