Communication, Overview about communication, Serial interface and printer – Flintec FT-13 User Manual

Page 34: Standard serial interface connector (rs232c), 1 overview about communication, 2 serial interface and printer, 1 standard serial interface connector (rs232c), Ommunication

Advertising
background image

FT-13 Technical Manual, Rev. 1.40 March 2013

Page 34 of 56

8. C

OMMUNICATION

8.1 Overview about Communication

Application

Hardware-Channel

Interface Operation mode

Data export to PC; Remote
display; Remote control via
ASCII input (Z, T, C)*

Standard Serial Interface (RS232C)

OR

Option Serial Interface 2 (RS232C, RS485, 20mA TTY)

OR

Option Ethernet (see Chap. 8.3)

Continuous output mode; see
Chap. 8.2.3 and Appendix 2

IndFace-Software

Host mode; see Chap. 8.2.4

Data export to PC; Output
to printer; Remote control
via ASCII input (P, Z, T, C)*

Print mode; see Chap. 8.2.5

Process control
(PLC, DCS, SCADA...)

Option Serial Interface 2 (RS485)

OR

Option Ethernet (see Chap. 8.3)

Modbus RTU;
see Appendix 3

Option Profibus; see Chap. 8.4 + Appendix 4

Profibus DP

Option ProfiNet; see Chap. 8.5 + Appendix 4

ProfiNet

Option CANopen, see Chap. 8.6 + Appendix 5

CANopen

* Z = Zeroing; T = Tara; C = Clear; P = Print

8.2 Serial Interface and Printer

8.2.1 Standard Serial Interface Connector (RS232C)

Standard 1

st

serial interface; Parameter group [00-]

Baudrate

1200, 2400, 4800, 9600, 19 200, 38 400 or 57 600 baud

Data format

8 bit no parity OR 7 bit even parity OR 7 bit odd parity

Start / Stop bit

1 start bit and 1stop bit

The connection of the standard RS232C of the instrument should be made as given in the table below.

Definition

Panel type housing;

Pin no. (D-Sub, 9-pin, male)

Stainless steel housing
Pin no. (J11 connector)

RS232C;

typical PC allocation

TXD

2

3

3

RXD

3

4

2

GND

7

2

5

Shield

Connector body

1

You have to connect each TXD (output) with RXD (input) at the other end.

8.2.2 Optional Serial Interface Connector (RS232C, RS485, 20 mA TTY)

An additional serial interface can be added in addition to the standard RS232C port.

Optional 2

nd

serial interface; Parameter group [02-]

Baudrate; Data format; Start / Stop bit

As standard serial interface

Hardware

RS232C / RS485 / 20 mA TTY current loop (field changeable);

20 mA TTY current loop not available with stainless steel housing
Up to 32 instruments can be connected with enabled RS485 port.

The pin configuration of the optional D-Sub 25-pin female connector is given below.

Parameter [012]

Pin no.

for desk and panel types

(D-Sub, 25-pin, female)

Pin no. for stainless steel housing

is set to

“0” or “1”

is set to “2”

J10 connector

(Option 1)

JR1 / JR2 connector

(Option 2 or 3)

TxD2; Par. [02-]

RTS1; Par. [01-]

4

3

10

RxD2; Par. [02-]

CTS1; Par. [01-]

5

2

9

GND (RS 232C)

7

9

13

R-; Par. [02-]

8

Not used

R+; Par. [02-]

9

Not used

T+; Par. [02-]

10

Not used

+V; Par. [02-]

11

Not used

Termination (RS 485)

19

Not used

Termination (RS 485)

20

Not used

A; Par. [02-]

21

5

12

B; Par. [02-]

22

4

11

T-; Par. [02-]

24

Not used

+24V; 20 mA CL

25

Not used

Shield

D25 body

1