Appendix 4: profibus dp / profinet data structure, Ppendix, Rofibus – Flintec FT-13 User Manual

Page 49: Rofi, Tructure, Ft-13 output

Advertising
background image

FT-13 Technical Manual, Rev. 1.40 March 2013

Page 49 of 56

A

PPENDIX

4:

P

ROFIBUS

DP

/

P

ROFI

N

ET

D

ATA

S

TRUCTURE

The output from FT-13 on PLC requests is in four words and according to the table below.

FT-13 Output

Bit no.

15

14

13

12

11

10

9

8

7

6

5

4

3

2

1

0

1

st

Word

W31

MSB

W30 W29 W28 W27 W26 W25 W24 W23 W22 W21 W20 W19 W18 W17

W16

2

nd

Word

W15

W14 W13 W12 W11 W10 W9

W8

W7

W6

W5

W4

W3

W2

W1

W0

LSB

3

rd

Word

0

0

0

0

In4

In3

In2

In1 Out8 Out7 Out6 Out5 Out4 Out3 Out2

Out1

4

th

Word

Error

0

0

0

0

0

0

0

0

Cmd

Acc.

Net

MT

Data

Ok

Data

Integrity


Here the first and the second word provide the weight information (32-bit signed integer; WXX = Weight bit XX;
W31 = MSD; W0 = LSD ) and the third and fourth word is the status as indicated below.

Bit no.

3

rd

Word

Description

15…12

Not used

Always 0

11

Digital Input 4

Input bit status (Active = 1)

10

Digital Input 3

9

Digital Input 2

8

Digital Input 1

7

Digital Output 8

Output bit status (Active = 1)

6

Digital Output 7

5

Digital Output 6

4

Digital Output 5

3

Digital Output 4

2

Digital Output 3

1

Digital Output 2

0

Digital Output 1

Bit no.

4

th

Word

Description in normal operation

15

Error

000

No error

001

ADC out

010

ADC over

011

ADC under

100

System error

101

FT not in weighing mode or filling mode

14

13

12…5

Not used

Always 0

4

Command accepted

This bit is toggled when a new command was accepted

3

Net

0: Gross Mode; 1: Net Mode

2

Motion

0: No Motion ; 1: Motion

1

Data Ok

0: Check the error bits; 1: Scale is operating properly

0

Update in progress
(Data integrity)

0: PLC should not use the data; 1: use the data
(if 0, the instrument was updating the PLC interface shared memory while data was
read. The PLC interface should ignore this data and rescan.)

Bit no.

4

th

Word

Description during the calibration

15 … 8

Calibration error

0000 0001

Calibration Timeout

– Restart calibration

0000 0010

ADC Error

– Re-energize the instrument

– If seen again, change board.

0000 0011

Instrument cannot be calibrated

– Check load cell cable

– Re-energize the instrument

0010 0010

Instrument cannot be calibrated

– Load cell signal is too low or too high

0010 0011

Calibration Error
– Calibration test weight is too small, increase calibration weight
– Check load cell connections

0010 0101

Scale unstable

– Wait until scale becomes stable

– Check grounding wiring

0010 0110

Calibration Jumper is not installed- Check Calibration jumper

7...0

Calibration status

0000 0001

Ready for calibration

0000 0011

Zero calibration in process

0000 0100

Span calibration in process

0000 1001

Calibration error (see bit 15...8)