The FT-13 Weighing Controller has been designed as very reliable and virtually error free instruments.
However if an error occurs do not attempt to repair the equipment before you understand what caused the
error. Note the problems you have with your instrument and the error messages shown on the display. Then try
to solve the problem according to the error table given below.

Error

Code

Description

Actions to take / Possible cause

Under

Weight is too low

Check the load;
Load cell or instrument could be defective

Over

Over Load

Adc Out

Input signal outside the input range

Check the load;
Check the calibration;
Load cell or instrument could be defective

E StbL

Stability error

- The scale must be stable for zeroing, taring, tolerance control, etc.
- Increase the maximum settling time (parameter [506])
- Increase the zeroing delay (parameter [50F])
- Increase the range of the motion detector (parameter [206])

Err 1

ADC error

Re-energize indicator;
Contact FLINTEC

Err 2
Err 3

Indicator cannot get calibrated

Check load cell and load; then repeat the calibration

Err 10

EEPROM error

Configure the instrument; EEPROM defective

Err 20

Calibration error

Calibrate the indicator

Err 21

Configuration error

Configure the indicator

Err 27

Indicator is not calibrated

Calibrate the indicator

Err 28

Date & Time Error

Adjust Date & Time;
Replace the battery

Err 30

Processor Error

Contact FLINTEC

Err 34

Indicator cannot get calibrated

Load cell signal is negative , very low or too high

Err 35

Calibration Error

Calibration test weight is too small;
Check load cell connections

Err 37

Scale unstable

Wait until scale becomes stable;
Check ground connection

Err 51

Filling programme was changed (previous
process parameters have been cleared)

Press the <Enter> key; then setup all parameters for the new
programme

Err 61

U10 (eeprom) is not installed

Contact FLINTEC

Err 62

U8 (eeprom) is not installed

E XXXX

Hardware error