Cleaning the analyzer – LABEC M-MIA-A User Manual
Page 20
 
Ultrasonic milk analyser
4. CLEANING THE ANALYZER
This procedure prevents gathering milk fat residues and milk stone on the 
sensor. The milk stone consists of milk solids, calcium, iron, sulphates, 
magnesium, etc. All these substances form layer on the pipe and sensor’s 
walls, which leads to deviations in the measurement results and blocking up 
the piping. 
 
The company-producer recommends usage of the chemicals, supplied with 
the analyser – alkaline and acidic (Lactodaily and Lactoweekly). You may 
order them separately or together with the analyser. Try to use only these 
chemicals for cleaning the analyser. 
 
4.1. Automatic cleaning the analyzer 
Analyzer's contamination as a result of the irregular cleaning is the basic 
reason for inexactness during measurement. In order to be avoid this, in the 
milk analyzers with peristaltic pump is embedded automatic cleaning. For this 
purpose the analyzer has to be prepared on the following way: 
Fig.6 Additional pipe to the input pipe for 
automatic cleaning 
94
 
Please, pay attention to the additional pipe, connected towards the input pipe. 
It has to be long enough to suck all the content in the glass, but to be bent in 
order not to impede the pump function. 
 
 
94 - additional pipe, 
connected towards 
input pipe 
Operation manual
MRC.VER.01-02.10