Immerse the electrode far enough into the solution to insure the reference 
junction is below the surface. 
Allow adequate time for the electrode to stabilize in standards and samples 
before taking a reading. 
Clean the electrode periodically. Allow more time for aged electrodes. 
Do not use the pH electrode in solutions of fluoride ion at low pH. This will 
etch the glass membrane. 
Sulphide vapors can permeate the electrode wick and contaminate the 
reference element. Minimize contact in such environments and change the 
reference electrolyte frequently. 
 
6. PH measuring. 
Measuring pH is an additional feature of the analyzer and is optional. 
Remove the protective rubber cap of the pH electrode. Take care to handle it 
appropriate in order not to be damaged. Use de-ionized or distilled water to 
rinse the electrode before usage. Fill in the sample holder with milk, put it in 
the recess of the analyzer and dip the pH electrode into the milk sample, 
ensuring complete dip of the electrode in the sample. Stir gently for 
homogenization of the sample. 
 
Measuring can be done in two modes: 
 
Off line by starting the menu pH & Co Meter | Measuring, when the 
analyzer works only as a pH meter. 
 
On line automatic pH measuring, when measuring the rest of the sample’s 
parameters. 
After starting the menu pH & Co Meter the following message appears on 
the display: 
 
 
pH Calibration 
pH Measuring 
pH En/Disable 
pH U Display 
------------------------------ 
Co Meter Calibr 
Co Meter Test 
Co Meter En/Dis 
Exit 
 